False-positive Histoplasma capsulatum Gen-Probe chemiluminescent test result caused by a Chrysosporium species.
J Clin Microbiol, 2005/3;43(3):1456-8.
Brandt ME[1], Gaunt D, Iqbal N, McClinton S, Hambleton S, Sigler L
Affiliations
PMID: 15750132
Impact factor: 11.677
Abstract
We describe a case in which the Histoplasma capsulatum AccuProbe test displayed cross-reactivity with a respiratory isolate thought to be Histoplasma but not morphologically consistent with H. capsulatum. The isolate was later identified as the Chrysosporium anamorph of Nannizziopsis vriesii by sequence analysis and phenotypic data.